IF MedTechBuild with us

Verification & validation

Proof that the device meets its requirements and works in real hands. Every protocol traces back to the risk file held under ISO 14971.

When verification and validation are the right step

When the design is stable enough to test against its requirements. Verification and validation turn the design file into evidence a regulator and an auditor can check.

  • Prove each requirement

    Bench tests that show each design output meets the design input it answers.

  • Show it works for users

    Design validation and summative usability testing with real users in realistic conditions.

  • Prepare for accredited labs

    Protocols, samples and pre-checks before safety, EMC or biocompatibility testing at an accredited lab.

What you bring, and what you get back

You bring

  • Design inputs and requirements
  • The risk management file
  • Design outputs: drawings, schematics and software
  • Units to test, built the way production will build them
  • Any test results so far

You get back

  • Verification and validation plan
  • Test protocols with acceptance criteria
  • Test reports
  • Accredited lab reports, where a standard needs them
  • Requirements trace matrix
  • Summative usability report

Verification and validation answer different questions

Design inputsRequirementseach with a numberVerificationDid we build it right?Output meets inputBench and lab testsValidationRight device for use?Meets user needsReal users and useOn production-equivalent units

Verification asks whether each design output meets its design input, usually on the bench. Validation asks whether the device meets the user's needs for its intended use, with real users in realistic conditions. ISO 13485 expects validation on initial production units or their equivalent, so units built by hand in the lab may not count. A device can pass every verification test and still fail validation.

How verification and validation work

  1. Step 1

    Plan

    A verification and validation plan traced to the requirements and the risks.

  2. Step 2

    Write the protocols

    Each test with its method, its sample and its acceptance criteria.

  3. Step 3

    Test

    In-house, and at accredited labs for the standards that need them.

  4. Step 4

    Report

    Reports and a trace matrix that shows every requirement covered.

Questions clients ask

When should verification and validation planning start?

When the requirements are written. A requirement that cannot be tested cannot be verified, so the plan checks each one early, while it can still be rewritten.

How many samples do we test?

The plan sets it for each test from the risk of failure and the confidence needed, and records why. Some standards also set a minimum.

Which tests need an accredited lab?

Tests to recognised standards, such as IEC 60601-1 electrical safety and IEC 60601-1-2 EMC, are usually run at a lab accredited to ISO/IEC 17025. Many other tests can run in-house under a controlled protocol.

What is a trace matrix?

A table that links each requirement to its design output, its risk controls and the test that proves it. It shows a reviewer that nothing was missed. See the design history file.

An assembly bench with tools and parts trays

Tested in-house and at accredited labs

Tests run in our own labs in Navi Mumbai, and at accredited labs for the standards that need them. Verification is one of nine disciplines in one team, so the test engineers work beside the designers.

Our quality system · About us

Devices we have worked on

We name the type of device, not the client. Client work stays confidential. See every device type.

Run under an ISO 13485:2016 quality system certified by TÜV SÜD, certificate 99 400 00046. What it covers.

Not quite you?

Standards this work is done against

  • ISO 14971Risk management
  • IEC 62366-1Usability engineering
  • IEC 60601 seriesElectrical safety and EMC, at accredited labs
  • ISO/IEC 17025Competence of the testing laboratories used

The list for a given device, and the edition of each, is set in its regulatory plan.

Where this sits

  1. 01Co-Innovate
  2. 02Co-Design & Develop
  3. 03Co-Manufacturing
  4. 04Co-Commercialize
  5. +Co-Invest

Also in Co-Design & Develop

Have a question about Verification & validation?

Opens the contact form with this service named. Nothing is sent until you press send.